Performance Tests

Performance Tests

44

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

3 Move to the Laser Source channel, move to the wavelength

parameter, [

λ], press Enter, select the lower wavelength source, and

press Enter.

4 If you are using an Agilent 81657A Laser Source make sure you

initialize the Agilent 8156A Optical Attenuator with 30 dB

attenuation.

5 Turn the instruments on, enable the laser source and allow the

instruments to warm up for at least 20 minutes.

6 Perform the following sub-procedure for both Power Meters:

a

Move to the Power Meter channel.

b

Move to the wavelength parameter, [

λ], press Enter, enter the

wavelength of the laser source, and press Enter.

c

Move to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.

d

Move to the averaging time parameter, [AvgTime], press Enter,
move to <500 ms>, and press Enter.

e

Move to the power parameter, [P], press [Pwr unit], move to
<Watt>, and press Enter.

7 Make sure the optical input of the Device Under Test (DUT),

8162xA/B, is not receiving any light by placing a plastic cap over the

input. Move to the DUT Power Meter channel, press [Menu], move to

<Zero>, press Enter.

8 Ensure that the Agilent 8156A output is disabled. Move to the

reference Power Meter channel that uses the Agilent 8162xB Optical

Head and Agilent 81618A Optical Head Interface module, press

[Menu], move to <Zero>, press Enter.

9 Enable the Agilent 8156A output and change the attenuation until

the reference Power Meter displays 10.00 mW.

10 Connect the attenuator output cable to the DUT Power Meter. Note

the power value returned from the DUT, [P], from the display and

note the result in the test record.

11 Move to the Laser Source channel, move to the wavelength

parameter, [

λ], press Enter, select the longer wavelength source,

and press Enter.

12 Repeat steps 4 to 10 at the second wavelength with the

corresponding source.