beautypg.com

Definition of terms, Averaging time, Linearity – Atec Agilent-81618A User Manual

Page 23

background image

Definition of Terms

Specifications

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

23

Definition of Terms

This section defines terms that are used both in this chapter and

“Performance Tests” on page 35.

Generally, all specifications apply for the given environmental

conditions and after warmup time.

Measurement principles are indicated. Alternative measurement

principles of equal value are also acceptable.

Averaging Time

Time defining the period during which the power meter takes readings

for averaging. At the end of the averaging time the average of the

readings is available (display- or memory-update). Symbol T

avg

.

Linearity

The linearity error is defined as the relative difference between the

displayed power ratio, D

x

/D

0

, and the actual (true)

power ratio P

x

/P

0

caused by changing the displayed power level from the reference level,

D

0

, to an arbitrary displayed level, D

x

. Symbol N.

if expressed in %

if expressed in dB

Conditions: reference level 10

µW, displayed power levels within the

specified range, zero less than specified time prior to measurement.

Note 1: ideally N = 0 %, respectively 0 dB.

Note 2: the power-dependent nonlinearity, N(P

x

), can alternatively be

expressed by the following formula:

where r(P) is the power-dependent responsivity (for a power meter,

the responsivity is defined as the ratio of displayed power to actual

input power).

N

D

x

D

0

P

x

P

0

-----------------

1

и

ш

ж

ц 100

=

N

dB

10

D

x

D

0

P

x

P

0

-----------------

и

ш

ж

ц

log

=

N P

x

( )

r P

x

( ) r P

0

( )

r P

0

( )

---------------------------------

=