Atec Agilent-81618A User Manual
Page 69
Performance Tests
Performance Tests
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
69
Performance Test for the Agilent 81623B
Page 3 of 3
Model Agilent 81623B Optical Head
Report No.________
Date ________________
Test
No.
Test Description
Min.
Spec.
Result Max.
Spec.
Measurement
Uncertainty
I
Accuracy Test
[
µW]
measured at _______ nm
(1310nm) Output Power
9.72
µW
_______
10.28
µW
measured at _______ nm
(1550nm) Output Power
9.72
µW
_______
10.28
µW
II
Linearity Test
For Calculations you may want to use the appropriate sheet
Range
P
DUT
[dBm]
P
DUT
[dBm]
Loss [%]
+10
+9
________
_______
<± 1.13 %
+10
+3
________
_______
<± 0.58 %
0
+3
________
_______
<± 0.58 %
0
-7
________
_______
<± 0.58 %
-10
-7
________
_______
<± 0.58 %
-10
-17
________
_______
<± 0.58 %
-20*
-17*
Reference
0.0
<± 0.58 %
-20
-27
________
_______
<± 0.58 %
-30
-27
________
_______
<± 0.58 %
-30
-37
________
_______
<± 0.63 %
-40
-37
________
_______
<± 0.63 %
-40
-47
________
_______
<± 1.08 %
-50
-47
________
_______
<± 1.08%
-50
-57
________
_______
<± 5.59 %
III
Noise Test
[pW]
_______
< 100 pW
IV
Return Loss Test
[dB]
50 dB <
_______