beautypg.com

Calculation, Example: measurement results – Atec Agilent-81618A User Manual

Page 51

background image

Performance Tests

Performance Tests

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

51

71 Increase the attenuation of Atty1 by 10 dB and note the results in

the test record (#20).

Calculation

72 Calculate the non-linearity using the formulas given in the test

record.

Example: Measurement Results

Information only

Your Entries

n

Atty1 / #221

[dB]

Atty2 / #100

[dB]

DUT Range

[dBm]

REF

Power [dBm]

DUT Power

[dBm]

Notes

1

0

42

10

-47,1915

9,4273

1. Value

2

6,6

42

10

-53,769

2,8467

InRange

3

6,6

35,5

10

-47,262

2,8524

RangeDisc / lower limit

4

6,6

35,5

0

-47,2605

2,8543

RangeDisc / upper limit

5

16,6

35,5

0

-57,2304

-7,1141

InRange

6

16,6

25,5

0

-47,2363

-7,1118

RangeDisc / lower limit

7

16,6

25,5

-10

-47,2347

-7,1097

RangeDisc / upper limit

8

26,6

25,5

-10

-57,2345

-17,111

InRange

9

26,6

15,5

-10

-47,2411

,17,106

RangeDisc / lower limit

10

26,6

15,5

-20

-47,2382

-17,1054

RangeDisc / upper limit

11

36,6

15,5

-20

-57,228

-27,0942

InRange

12

36,6

5,5

-20

-47,2317

-27,0864

RangeDisc / lower limit

13

36,6

5,5

-30

-47,2318

-27,0873

RangeDisc / upper limit

14

46,6

5,5

-30

-57,2336

-37,089

InRange

15

35,5

17,6

-30

-47,219

-37,253

RangeDisc / lower limit

16

35,5

17,6

-40

-47,2195

-37,22538

RangeDisc / upper limit

17

45,5

17,6

-40

-57,1333

-47,166

InRange

18

45,5

7,6

-40

-47,1411

-47,164

RangeDisc / lower limit

19

41,2

11,3

-50

-47,1355

-47,1664

RangeDisc / upper limit

20

51,2

11,3

-50

-57,134

-57,163

InRange