Performance Tests
Performance Tests
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
51
71 Increase the attenuation of Atty1 by 10 dB and note the results in
the test record (#20).
Calculation
72 Calculate the non-linearity using the formulas given in the test
record.
Example: Measurement Results
Information only
Your Entries
n
Atty1 / #221
[dB]
Atty2 / #100
[dB]
DUT Range
[dBm]
REF
Power [dBm]
DUT Power
[dBm]
Notes
1
0
42
10
-47,1915
9,4273
1. Value
2
6,6
42
10
-53,769
2,8467
InRange
3
6,6
35,5
10
-47,262
2,8524
RangeDisc / lower limit
4
6,6
35,5
0
-47,2605
2,8543
RangeDisc / upper limit
5
16,6
35,5
0
-57,2304
-7,1141
InRange
6
16,6
25,5
0
-47,2363
-7,1118
RangeDisc / lower limit
7
16,6
25,5
-10
-47,2347
-7,1097
RangeDisc / upper limit
8
26,6
25,5
-10
-57,2345
-17,111
InRange
9
26,6
15,5
-10
-47,2411
,17,106
RangeDisc / lower limit
10
26,6
15,5
-20
-47,2382
-17,1054
RangeDisc / upper limit
11
36,6
15,5
-20
-57,228
-27,0942
InRange
12
36,6
5,5
-20
-47,2317
-27,0864
RangeDisc / lower limit
13
36,6
5,5
-30
-47,2318
-27,0873
RangeDisc / upper limit
14
46,6
5,5
-30
-57,2336
-37,089
InRange
15
35,5
17,6
-30
-47,219
-37,253
RangeDisc / lower limit
16
35,5
17,6
-40
-47,2195
-37,22538
RangeDisc / upper limit
17
45,5
17,6
-40
-57,1333
-47,166
InRange
18
45,5
7,6
-40
-47,1411
-47,164
RangeDisc / lower limit
19
41,2
11,3
-50
-47,1355
-47,1664
RangeDisc / upper limit
20
51,2
11,3
-50
-57,134
-57,163
InRange