Safety Considerations

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

3

Safety Considerations

The following general safety precautions must be observed during all

phases of operation, service, and repair of this instrument. Failure to

comply with these precautions or with specific warnings elsewhere in

this manual violates safety standards of design, manufacture, and

intended use of the instrument. Agilent Technologies Inc. assumes no

liability for the customer’s failure to comply with these requirements.

Before operation, review the instrument and manual, including the

red safety page, for safety markings and instructions. You must follow

these to ensure safe operation and to maintain the instrument in safe

condition.

W A R N I N G

The WARNING sign denotes a hazard. It calls attention to a procedure,

practice or the like, which, if not correctly performed or adhered to,

could result in injury or loss of life. Do not proceed beyond a

WARNING sign until the indicated conditions are fully understood and

met.

Safety Symbols

The apparatus will be marked with this symbol when it is necessary

for the user to refer to the instruction manual in order to protect the

apparatus against damage.

Caution - hot surface.

Magnetic fields - may interfere with a pacemaker.

Input Power Limitations

W A R N I N G

Applicable to Agilent 81628B only -

Operation above 34 dBm (2.51 W) is at your own risk.

Operation at a continuous optical power, or an average optical power,

greater than 34 dBm causes, in the specified operating temperature

range, the metal parts (connector adapters, screws) to become hotter

than the required limit (55°C) of the safety standard IEC 61010-1.