Performance Tests

Performance Tests

48

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

0 dBm Range (equivalent to +30 dBm range for 81628B)

18 Disable Atty1.

19 Zero both Power Meters. On the 8163A/B with two installed power

meters, press [Menu], move to <Zero all>, and press Enter.

20 Enable Atty1.

21 Switch one range up to the +10 dBm range (+40dBm for 81628).

22 Note both power readings (#3).

23 Switch down to the previous range (0 dBm, 30 dBm for the 81628B)

and note the values again (#4).

24 Increase Atty1 by 10 dB and note the results in the test record (#5).

25 At the DUT, 8162xB, press [Menu], move to <Range>, press Enter,

move to <-10 dBm>(20 dBm for the 81628B), press Enter, and press

[Close].

26 If necessary, adjust the attenuation of Atty1 in order to be on the

upper limit of the range (i.e. -x7.y dBm).

27 Decrease the attenuation of Atty2 by 10 dB in order to be on the

upper limit of the -50 dBm range at the reference power meter.

-10 dBm Range (equivalent to +20 dBm range for
81628B)

28 Disable Atty1

29 Zero both Power Meters. On the 8163A/B with two installed power

meters, press [Menu], move to <Zero all>, and press Enter.

30 Enable Atty1.

31 Switch one range up to the 0 dBm range (30 dBm for the 81628B)

and note the power readings (#6).

32 Switch down to the previous range (-10 dBm, 20 dBm for the

81628B) and note the values again (#7).

33 Increase the attenuation of Atty1 by 10 dB and note the results in

the test record (#8).

34 At the DUT, 8162xB, press [Menu], move to , press Enter,

move to <-20 dBm>(10 dBm for the 81628B), press Enter and press

[Close].

35 If necessary, adjust the attenuation of Atty1 in order to be on the

upper limit of the range (i.e. -x7.y dBm).