beautypg.com

Atec Agilent-81618A User Manual

Page 84

background image

Performance Tests

Performance Tests

84

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

Performance Test for the Agilent 81628B

Page 3 of 3

Model Agilent 81628B Optical Head

Report No.________

Date ________________

Test
No.

Test Description

Min.
Spec.

Result Max.

Spec.

Measurement

Uncertainty

I

Linearity Test (without sphere)

For Calculations you may want to use the appropriate sheet

Range

P

DUT

[dBm]

P

DUT

[dBm]

Loss [%]

+40

+9

________

_______

<± 0.46 %

+40

+3

________

_______

<± 0.46 %

+30

+3

________

_______

<± 0.46 %

+30

-7

________

_______

<± 0.46 %

+20

-7

________

_______

<± 0.46 %

+20

-17

________

_______

<± 0.46 %

+10*

-17*

Reference

0.0

<± 0.46 %

+10

-27

________

_______

<± 0.46 %

0

-27

________

_______

<± 0.46 %

0

-37

________

_______

<± 0.46 %

-10

-37

________

_______

<± 0.46 %

-10

-47

________

_______

<± 0.49 %

-20

-47

________

_______

<± 0.49 %

-20

-57

________

_______

<± 0.71 %

II

Accuracy Test (with sphere)

[

µW]

measured at _______ nm
(1310nm) Output Power

9.64

µW

_______

10.36

µW

measured at _______ nm
(1550nm) Output Power

9.64

µW

_______

10.36

µW

III

Noise Test

[pW]

_______

< 5 nW