Performance Tests

Performance Tests

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

65

[1] The 20 dBm range measurement is only performed to check the

functionality of the current booster. The published specification does

not apply because the attenuator is not in automode.

Performance Test for the Agilent 81622B

Page 3 of 3

Model Agilent 81622B Optical Head

Report No.________

Date ________________

Test
No.

Test Description

Min.
Spec.

Result Max.

Spec.

Measurement
Uncertainty

I

Accuracy Test

[

µW]

measured at _______ nm
(1310nm) Output Power

9.64

µW

_______

10.36

µW

measured at _______ nm
(1550nm) Output Power

9.64

µW

_______

10.36

µW

II

Linearity Test

For Calculations you may want to use the appropriate sheet

Range

P

DUT

[dBm]

P

DUT

[dBm]

Loss [%]

+20

+9

<± 1.53

[1]

%

+10

+9

________

_______

<± 1.16 %

+10

+3

________

_______

<± 1.16 %

0

+3

________

_______

<± 1.16 %

0

-7

________

_______

<± 1.16 %

-10

-7

________

_______

<± 1.16 %

-10

-17

________

_______

<± 1.36 %

-20*

-17*

Reference

0.0

<± 1.36%

-20

-27

________

_______

<± 3.16 %

-30

-27

________

_______

<± 3.16 %

-30

-37

________

_______

<± 21.21 %

III

Noise Test

[nW]

_______

< 40 nW

IV

Return Loss Test

[dB]

45 dB <

_______