Atec Agilent-81618A User Manual
Page 65
Performance Tests
Performance Tests
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
65
[1] The 20 dBm range measurement is only performed to check the
functionality of the current booster. The published specification does
not apply because the attenuator is not in automode.
Performance Test for the Agilent 81622B
Page 3 of 3
Model Agilent 81622B Optical Head
Report No.________
Date ________________
Test
No.
Test Description
Min.
Spec.
Result Max.
Spec.
Measurement
Uncertainty
I
Accuracy Test
[
µW]
measured at _______ nm
(1310nm) Output Power
9.64
µW
_______
10.36
µW
measured at _______ nm
(1550nm) Output Power
9.64
µW
_______
10.36
µW
II
Linearity Test
For Calculations you may want to use the appropriate sheet
Range
P
DUT
[dBm]
P
DUT
[dBm]
Loss [%]
+20
+9
<± 1.53
[1]
%
+10
+9
________
_______
<± 1.16 %
+10
+3
________
_______
<± 1.16 %
0
+3
________
_______
<± 1.16 %
0
-7
________
_______
<± 1.16 %
-10
-7
________
_______
<± 1.16 %
-10
-17
________
_______
<± 1.36 %
-20*
-17*
Reference
0.0
<± 1.36%
-20
-27
________
_______
<± 3.16 %
-30
-27
________
_______
<± 3.16 %
-30
-37
________
_______
<± 21.21 %
III
Noise Test
[nW]
_______
< 40 nW
IV
Return Loss Test
[dB]
45 dB <
_______