beautypg.com

Dh-21 – Xylem DH-21 User Manual

Page 68

background image

4-22 Using the Graphical “W

ATER

LOG” Software

DH-21

SCAN RATE

HYPER DIVISOR

FAST SCAN RATE

1 minute

60

1 seconds

1 minute

30

2 seconds

1 minute

20

3 seconds

1 minute

15

4 seconds

1 minute

12

5 seconds

1 minute

10

6 seconds

1 minute

6

10 seconds

1 minute

5

12 seconds

1 minute

4

15 seconds

1 minute

3

20 seconds

1 minute

2

30 seconds

1 minute

1

60 seconds

This option allows the sensor to capture and record quick changing data without the expense of
battery life by adjusting the scan rate based on the data change rate. It also saves memory by not
logging data when the data is not changing significantly.

Note that this check is on the level parameter only, not the temperature, and the values entered
for the hyper limit and delta limit will be in the same units as set in the coefficients options. For
example, if the coefficients Pressure / Level option is set to feet, the hyper limit and delta limit
will also be in feet.