Using the Graphical “W

ATER

LOG” Software 4-57

In the stand-alone mode the “

W

ATER

LOG” software will not display all status information of the

sensor and it will not perform functions requiring the sensor to be connected. These functions
include setting the time and date, forcing the sensor to make a measurement, zeroing the sensor,
extracting data, setting the operation mode and sensor diagnostics.

The stand-alone mode is handy for setting up the majority of the options for the sensor while not
connected or attached. The setups can then be saved to a disk file. It is important to save any
desired settings to a file before attaching to the sensor because the setups in the sensor will
overwrite the setups in the PC “

W

ATER

LOG” software when it is attached. If the sensor is

attached and a setup file is retrieved, the settings in the setup file will be transferred to the sensor.

Figure 4-39. Attach to Sensor Option