3-4 Using the “W
ATER
LOG” Software
DH-21
Warning : This option should not be invoked if the user has uncollected data in the
FLASH EEPROM which needs to be extracted. In the event of a dead battery, the memory pointers
in the “
W
ATER
LOG” will point to the last location in memory. Therefore, you can use the extract
data menu option number “8” to extract the data before resetting the “
W
ATER
LOG”.
3.2 Configuration Options Menu
3.2.1 Reset Option
For now we will skip the Run and Off options. Figure 3-2 shows a sample screen when the
“
W
ATER
LOG” sensor is first installed. Notice the time and date values are meaningless and
need to be initialized. The first option that should be invoked is the Reset (option 3). Press the
number “3” key followed by
sensor will be reset. Enter a “y”
The WaterLOG Serial # T2000 Version 1.1
Status = OFF (LINEAR) Last Measurement Units
Date/Time = 01/01/95 12:00:47
Date Started = 01/01/95 14:39:00 PRESSURE = 0.0000 PSI
Logging Rate = 00:01:00 TEMPERATURE = 0.00 Deg C
Next Log Time = 00:00:00
Memory Used = 100.0 %
Logging Count = 10921 Scans
Battery = 6.8 Volts
1) Run 4) Time/Date 7) User Notes 10) Measure Data
2) Off 5) Logging Mode 8) Extract Data 11) Zero Sensor
3) Reset 6) Coefficients 9) View Data 12) Sensor Info
13) Diagnostics
Enter Option: 3 Esc-Exit
Reset
This will erase all data and reset the sensor.
Continue? y (y/n)
Esc-Exit
Figure 3-2. Reset Option