Dh-21 – Xylem DH-21 User Manual

Page 43

background image

DH-21

Using the “W

ATER

LOG” Software 3-25

The WaterLOG Serial # T2000 Version 1.1

Status = OFF (VARIABLE) Last Measurement Units
Date/Time = 02/28/95 04:12:23
Date Started = 02/28/95 04:00:00 PRESSURE = -0.0051 PSI
Logging Rate = 00:01:00 TEMPERATURE = 78.38 Deg F
Next Log Time = 04:12:51
Memory Used = 0.1 %
Logging Count = 16 Scans
Battery = 6.8 Volts


1) Run 4) Time/Date 7) User Notes 10) Measure Data
2) Off 5) Logging Mode 8) Extract Data 11) Zero Sensor
3) Reset 6) Coefficients 9) View Data 12) Sensor Info
13) Diagnostics
Enter Option: 12 Esc-Exit

Current WaterLOG Sensor Information
Vendor = Design Analysis Associates
Model # = WaterLOG Firmware Version = V11 PCB Serial # = 000000
Serial # = T2000 Product Revision = 00A
Type any key to continue. Esc-Exit

Figure 3-22. Read Sensor Information Option

3.2.13 Diagnostics Option

The Diagnostics option, as shown in Figure 3-23, performs several tests on the hardware of the

W

ATER

LOG” sensor. This option is used to confirm that the sensor is working properly. The

sensor must be in the reset mode before this option can be selected.

Note: Time, Date, User Notes and ALL DATA will be overwritten during this test.

Extract any desired data before using this option.