beautypg.com

Dh-21 – Xylem DH-21 User Manual

Page 34

background image

3-16 Using the “W

ATER

LOG” Software

DH-21

The “

W

ATER

LOG” stores 160 characters of User Notes in the FLASH EEPROM. These notes

allow you to enter information which will be recovered during data extraction. Figure 3-11
shows a sample User Notes screen. The cursor is placed in the “User Notes Edit/View” portion
of the screen when option 7 is invoked. You can then overwrite or edit the existing notes. To
store the entered User Notes, simply press the key. This means the key
should not be pressed to go from line to line. The space bar can be used to complete a line where
the cursor will wrap around to the next line. You can also use the UP , Down and LEFT ,
RIGHT arrow keys to move the cursor around the User Notes. When the key is
pressed, the software in the computer will down-load the “

W

ATER

LOG” with the notes. This

process takes several seconds to complete. It might appear nothing is happening because of the
time required to transfer the data. The software will return to the main menu when the data has
been transferred.

The WaterLOG Serial # T2000 Version 1.1

Status = RESET (VARIABLE) Last Measurement Units
Date/Time = 02/28/95 03:56:29
Date Started = 01/01/95 14:39:00 PRESSURE = 0.0000 PSI
Logging Rate = 00:00:05 TEMPERATURE = 0.00 Deg F
Next Log Time = 00:00:00
Memory Used = 0.0 %
Logging Count = 0 Scans
Battery = 6.8 Volts


1) Run 4) Time/Date 7) User Notes 10) Measure Data
2) Off 5) Logging Mode 8) Extract Data 11) Zero Sensor
3) Reset 6) Coefficients 9) View Data 12) Sensor Info
13) Diagnostics
Enter Option: 7 Esc-Exit

User Notes Edit/View

WaterLOG Down Hole Data Logger
From
Design Analysis Associates, Inc.
Logan, Utah 84321 (801) 753-2212

Figure 3-11. User Notes Option

The User Notes can be used to store information while the sensor is logging. This is useful for
entering site conditions, for example: “Jan 1, 96 Replaced Battery”.