3-26 Using the “W
ATER
LOG” Software
DH-21
The WaterLOG Serial # T2000 Version 1.1
Status = RESET (VARIABLE) Last Measurement Units
Date/Time = 02/28/95 04:13:02
Date Started = 02/28/95 04:00:00 PRESSURE = -0.0051 PSI
Logging Rate = 00:01:00 TEMPERATURE = 78.38 Deg F
Next Log Time = 04:12:51
Memory Used = 0.0 %
Logging Count = 0 Scans
Battery = 6.8 Volts
1) Run 4) Time/Date 7) User Notes 10) Measure Data
2) Off 5) Logging Mode 8) Extract Data 11) Zero Sensor
3) Reset 6) Coefficients 9) View Data 12) Sensor Info
13) Diagnostics
Enter Option: 13 Esc-Exit
Diagnostics
This hardware diagnostic overwrites
time, date, user notes, and logged data.
Continue? y (y/n)
Figure 3-23. Diagnostics Option
Figure 3-24 and Figure 3-25 show the results of the diagnostics. The first line tests the CONFIG
register to see if the processor's COP timer is enabled. The diagnostics software in the
“
W
ATER
LOG” then checks the program EEPROM memory for a correct checksum. The
software then tests the RAM Memory, FLASH EEPROM memory, EEPROM user notes, and
performs a measurement that returns pressure, raw pressure voltage, temperature, and raw
temperature voltage. The results of the tests will be shown on the screen. After the tests have
completed, you can press any key to return to the main menu.
WaterLOG Self Diagnostics
CONFIG Register = 09: PASSED
Starting EPROM checksum test
EPROM checksum test: PASSED
Starting Clock Test
Set New Time: 23:59:59
Wakeup Time: 00:00:04
WaterLOG is going to sleep for 5 seconds
WaterLOG is now awake: PASSED
Starting RAM test
Pattern test a RAM location to verify data bus
Load odd-length pattern into RAM to verify address bus
Verify odd-length pattern in RAM
RAM test: PASSED
Starting FLASH test
Load odd-length pattern into FLASH to verify address bus
Verify odd-length pattern in FLASH
Load zeros into FLASH
Verify zeros in FLASH
Esc-EXIT
Figure 3-24. Self-Test Diagnostic Display