3-6 Using the “W

ATER

LOG” Software

DH-21

3.2.3 Logging Mode Option

Options 5, 6, and 7 configure the “

W

ATER

LOG” for unattended logging. Option 5 is shown in

Figure 3-4. The logging rate indicates the elapsed time between measurements. The

W

ATER

LOG” clock circuitry allows the “

W

ATER

LOG” to enter a low-power “SLEEP” mode

of operation between measurements. Then when the selected measurement time interval has
elapsed, which is specified in option 5, the “

W

ATER

LOG” will “WAKE-UP” and take a

pressure, temperature and battery measurement. It will then store the data in FLASH EEPROM
with a time-stamp, and go to “SLEEP” again.

The time between samples can range from 1 second up to 24 hours. If the logging rate is less
than 6 seconds, the “

W

ATER

LOG” will not go to “SLEEP”.

The “

W

ATER

LOG” “WAKES-UP” and takes a measurement when the time in the DATE/TIME

field matches the “Next Log Time”. The “Next Log Time” is incremented by the logging rate. If
the new “Next Log Time” has already occurred, the “

W

ATER

LOG” will continue to add the

logging rate to the new Next Log Time until the new Next Log Time is in the future. Then the

W

ATER

LOG” will “WAKE-UP” to take the next sample.

The WaterLOG Serial # T2000 Version 1.1

Status = RESET (LINEAR) Last Measurement Units
Date/Time = 02/28/95 03:50:24
Date Started = 01/01/95 14:39:00 PRESSURE = 0.0000 PSI
Logging Rate = 00:01:00 TEMPERATURE = 0.00 Deg C
Next Log Time = 00:00:00
Memory Used = 0.0 %
Logging Count = 0 Scans
Battery = 6.8 Volts


1) Run 4) Time/Date 7) User Notes 10) Measure Data
2) Off 5) Logging Mode 8) Extract Data 11) Zero Sensor
3) Reset 6) Coefficients 9) View Data 12) Sensor Info
13) Diagnostics
Enter Option: 5 Esc-Exit

Logging Mode
1) Linear Mode 4) Hyper-Delta Mode
2) Delta Mode 5) Variable Mode
3) Hyper Mode
Enter Option: 1 Esc-Exit

Figure 3-4. Logging Rate Option