beautypg.com

To set modulating amplitude, To set modulating polarity – RIGOL DG1000Z Series User Manual

Page 97

background image

Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-47

To Set Modulating Amplitude

Press ASK Ampl to set the modulating amplitude。
 Use the numeric keyboard or direction keys and knob to input the desired

amplitude value.

 The range of amplitude (HighZ) is from 0 to 10Vpp and the default is 2Vpp.

To Set Modulating Polarity

Press Polarity to select the “Pos” or “Neg” polarity of the modulating waveform to

control the output amplitude.

In internal modulation, set the polarity to “Pos” and the generator would output the

lower of the carrier amplitude and modulating amplitude when the modulating

waveform is logic low level and output the greater when the modulating waveform is

logic high level. The situation is the opposite when the polarity is set to “Neg”.

In external modulation, set the polarity to “Pos” and the generator would output the

lower of the carrier amplitude and modulating amplitude when the external input

signal is logic low level and output the greater when the external input signal is logic

high level. The situation is the opposite when the polarity is set to “Neg”.。