Chapter 4 Troubleshooting

4-2

DG1000Z User’s Guide

3) Restarting the instrument and insert the USB storage device again to check

whether it can work normally.

4) If the USB storage device still can not be used normally, please contact

RIGOL.


6. How to set the amplitude of the waveform in dBm?

1) Press CH1|CH2 to select the desired channel.
2) Press UtilityChannel SetOutput SetImped to check whether

the setting is “HighZ”. If yes, you can not set the amplitude of the waveform

in dBm at this time. Press Imped to select “Load” and use the numeric

keyboard or direction keys and knob to set the impedance to a proper

value.

3) Select the desired waveform, press Ampl/HiLevel to highlight “Ampl”, and

then input the desired value using the numeric keyboard. Then select the

unit “dBm” from the pop-up menu.

7. Performance verification test is not passed:

1) Check whether the generator is within calibration period (1 year).

2) Check whether the generator is warmed up for at least 30 minutes before

the test.

3) Check whether the generator is

under the specified temperature.

4) Check whether the test is under strong-magnetism environment.

5) Check whether the power supplies of the generator and test system have

strong interference.

6) Check whether the performance of the test device used meets the

requirement.

7) Make sure that the test device used is within the calibration period.

8) Check whether the test devices used meets the required conditions of the

manual.

9) Check whether all the connections are tight.

10) Check whether any cable has internal damage.

11) Make sure that the operations conform to settings and processes which are

required by the performance verification manual.

12) Check whether the error calculation exits faults.

13) Correctly understand the definition of “typical value” for this product: the

performance specification of this product under specified conditions.