Counter

DG1000Z provides a counter which can measure various parameters (such as

frequency, period, duty cycle, positive pulse width and negative pulse width) of

external input signal and support statistic of measurement results. When the statistic

function is enabled, the instrument calculates the maximum, minimum, average and

standard deviation of the measurement values automatically and displays the

variation tendency of the measurement values in “Digital” or “Curve” mode. In

addition, the dual channels can still output normally when the counter is enabled.

To Enable the Counter


Press Counter at the front panel (the backlight turns on and the left indicator
flickers) to enable counter function and enter counter setting interface.

Figure 2-13 Counter Parameter Setting Interface

If the counter is currently turned on and the screen displays the counter interface,

press Counter again to disable counter function. If the counter is currently turned
on and the screen displays interfaces other than the counter interface, press

Counter again to switch to the counter interface.

Note:

when the counter is enabled, the CH2 sync output will be disabled.

Coupling Trigger Level HF Suppression

Gate Time

Trigger

Sensitivity