RIGOL DG1000Z Series User Manual
Page 156
![background image](https://www.manualsdir.com/files/859125/content/doc156.png)
RIGOL
Chapter 2 Front Panel Operations
2-106
DG1000Z User’s Guide
To Set Subnet Mask
In ManualIP mode, the subnet mask could be manually set.
The format of subnet mask is nnn.nnn.nnn.nnn, wherein nnn ranges from 0 to
255. It is recommended that users acquire an available subnet mask from their
network administrator.
Press SubMask and use the direction keys and numeric keyboard or knob to
enter your desired subnet mask. The setting is stored in non-volatile memory
and will be automatically loaded when the generator is powered on the next
time if DHCP and AutoIP are set as “Off”.
To Set Default Gateway
In ManualIP mode, the default gateway could be manually set.
The format of default gateway is nnn.nnn.nnn.nnn, wherein the first nnn ranges
from 1 to 223 (except 127) and the other three nnn range from 0 to 255.
It is
recommended that users acquire an available default gateway from their
network administrator.
Press Default Gateway and use the direction keys and numeric keyboard or
knob to enter your desired gateway address. The setting is stored in non-volatile
memory and will be automatically loaded when the generator is powered on the
next time if DHCP and AutoIP are set as “Off”.
To Set DNS Service
In ManualIP mode, the DNS could be manually set.
The DNS format is nnn.nnn.nnn.nnn, wherein the first nnn ranges from 1 to 223
(except 127) and the other three nnn range from 0 to 255.
It is recommended
that users acquire an available DNS address from their network administrator.
Press DNS Service and use the direction keys and numeric keyboard or knob to
enter your desired DNS address. The setting is stored in non-volatile memory
and will be automatically loaded when the generator is powered on the next
time if DHCP and AutoIP are set as “Off”.
Default Configuration
Press Default Config and the prompt message “Restore network settings to preset
values of LXI?” is displayed. Press OK to restore the network parameters to their
default values. By default, DHCP and AutoIP are enabled and ManualIP is disabled.
Current Configuration
Press Current Config to examine the MAC address of the current instrument, the
current LAN parameters and LAN status information.