Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-63

3. Manual Trigger

A sweep will be generated from the corresponding channel once you press

Trigger when manual trigger is selected.

Note:

the menu key Trigger is valid when the “Manual Trigger” is selected and

the output of the corresponding channel is enabled.

Press TrigOut to select “Leading” or “Trailing” and the [CH1/Sync/Ext

Mod/Trig/FSK] connector at the rear panel will output trigger signal with

specified edge.

KEY POINTS:

1. When “Int” or “Manual” trigger is selected, press SweepTrigger

TrigOut to select “Off”, “Leading” or “Trailing”.

1) When “Off” is selected, the [CH1/Sync/Ext Mod/Trig/FSK]

connector at the rear panel will ouput the sync signal which changes from

low level to high level at the start of each sweep and returns to low level

at center frequency point or at the specified mark frequency point.

2) When “Leading” is selected, the [CH1/Sync/Ext Mod/Trig/FSK]

connector at the rear panel will ouput the trigger signal which is the same

as the sync signal. The trigger signal changes from low level to high level

at the start of each sweep (that is, enable the trigger at the leading edge)

and returns to low level at center frequency point or at the specified mark

frequency point.

3) When “Trailing” is selected, the [CH1/Sync/Ext Mod/Trig/FSK]

connector at the rear panel will ouput the trigger signal which changes

from high level to low level at the start of each sweep (that is, enable the

trigger at the trailing edge) and returns to high level at center frequency

point or at the specified mark frequency point.

[CH1/Sync/Ext Mod/Trig/FSK]

[CH2/Sync/Ext Mod/Trig/FSK]