Chapter 2 Front Panel Operations

2-60

DG1000Z User’s Guide

Figure 2-8 Log Sweep

Step Sweep

In Step Sweep type, the output frequency of the instrument “steps” from “Start

Frequency” to “Stop Frequency”. The duration of the output signal on each frequency

point is determined by “Sweep Time” and “Step Number”.


When Sweep is enabled, press Type to select “Step”. A step waveform is displayed
on the waveform on the screen, indicating that the output frequency varies in “step”.

At this point, press

to open the 2/2 menu page. Press Step Num and input the

step number using the numeric keyboard or direction keys and knob. The default is 2

and the range is from 2 to 1024.

Note:

in “Linear” and “Log” sweep modes, Step Num is grayed out and disabled.

Figure 2-9 Step Sweep

Sweep Time

When Sweep is enabled, press SweepTime and use the numeric keyboard or knob
to modify the sweep time. The default value is 1s and the available range is from 1ms

to 500s. The generator will renewedly sweep and output from the specified “Start

Frequency” once the sweep time is modified.

Return Time

Return time describes the time in which the output returns to “Start frequency” from

“Stop Frequency” after the generator has swept from “Start Frequency” to “Stop

Frequency” and stay within “Stop Hold” time.


When Sweep is enabled, press ReturnTime and use the numeric keyboard or
direction keys and knob to modify the return time. The default value is 0s and the

available range is from 0s to 500s.