Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-61

The generator will renewedly sweep from the specified “Start Frequency” once the

return time is modified.

Mark Frequency

The sync signal output from the [CH1/Sync/Ext Mod/Trig/FSK] connector

(corresponds to the channel at the front panel) at the rear panel always changes

from low level to high level at the start of each sweep. The sync signal changes from

high level to low level at the center frequency point when “Mark” function is disabled

or at the specified mark frequency point when “Mark” function is enabled.

When Sweep is enabled, press Mark to select “On” and use the numeric keyboard
or direction keys and knob to modify the mark frequency. The default value is 550Hz

and the available range is limited by “Start Frequency” and “Stop Frequency”.

The generator will renewedly sweep and output from the specified “Start Frequency”

once the mark frequency is modified.

KEY POINTS:

For step sweep (the sweep points determined by the start frequency, stop

frequency and step number respectively are f

1

, f

2

, ……, f

n

, f

n+1

……

), if the mark

frequency setting is one of the sweep points values, the sync signal is TTL high

level at the start of the sweep and will change to low level at the mark frequency

point. If the mark frequency setting is not any one of the sweep points values, the

sync signal will change to low level at the sweep point which is closest to this mark

frequency.

Start Hold

Start hold is the duration that the output signal outputs with the “Start Frequency”

after the sweep starts. After the start hold, the generator will output with varying

frequency according to the current sweep type.


When Sweep is enabled, press Start Hold and use the numeric keyboard or
direction keys and knob to change the Start Hold. The default value is 0s and the

available range is from 0s to 500s.


The generator will renewedly sweep and output from the specified “Start Frequency”
once the Start Hold is changed.