To select modulation source, To set modulating rate – RIGOL DG1000Z Series User Manual

Page 96

background image

RIGOL

Chapter 2 Front Panel Operations

2-46

DG1000Z User’s Guide

To Select Modulation Source

Press ModSource to select “Int” or “Ext” modulation source.

1. Internal Source

When internal source is selected, the modulating waveform is set as a Square

with 50% duty cycle, and the frequency at which the output amplitude “shifts”

between “carrier amplitude” and “modulating amplitude” is determined by “ASK

Rate”.

2. External Source

When external source is selected, the generator accepts the external modulating

signal from the [CH1/Sync/Ext Mod/Trig/FSK] connector at the rear panel.

Note:

for the

[CH1/Sync/Ext Mod/Trig/FSK] connector, controlling ASK

modulation externally is different from controlling AM/FM/PM modulations

externally. In ASK modulation, you can set the modulating polarity.

To Set Modulating Rate

When internal source is selected, press ASK Rate to set the rate at which the output

amplitude shifts between “carrier amplitude” and “modulating amplitude”.
 Use the numeric keyboard or direction keys and knob to input the desired

frequency value.

 The frequency range is from 2mHz to 1MHz and the default is 100Hz.

Note:

when external modulation source is selected, this menu will be grayed out and

disabled.

[CH1/Sync/Ext Mod/Trig/FSK]

[CH2/Sync/Ext Mod/Trig/FSK]