Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-43

To Select Modulation Source

DG1000Z can accept modulating waveform from internal or external modulation
source. Press ModSource to select “Int” or “Ext” modulation source.

1. Internal Source

When internal modulation source is selected, press Shape to select Sine, Square,

Triangle, UpRamp, DnRamp, Noise or Arb as modulating waveform. The default

is Sine.
 Square: 50% duty cycle
 Triangle: 50% symmetry
 UpRamp: 100% symmetry
 DnRamp: 0% symmetry
 Arb: the arbitrary waveform selected of the current channel

Note:

noise can be used as modulating waveform but can not be used as carrier

waveform.

2. External Source

When external modulation source is selected, Shape and PM Freq will be

grayed out and disabled. The generator accepts the external modulating signal

from the [CH1/Sync/Ext Mod/Trig/FSK] connector at the rear panel. At this

time, phase deviation of the modulated waveform is controlled by the ±5V signal

level of the connector. For example, if the phase deviation is set to 180°, +5V

signal level corresponds to a 180º phase variation. The lower the external signal

level, the less deviation would be generated.

[CH1/Sync/Ext Mod/Trig/FSK]

[CH2/Sync/Ext Mod/Trig/FSK]