Stat, Auto – AMETEK PLA-PLW Programming Manual User Manual
Page 69

65
STAT
E
AUTO
MODe:CURRent
STEP:CURRent[:LEVel] 1,15
; Program the STEP of input values for each
function. The STEP commands take a
comma-separated STEP of arguments.
STEP:CURRent[:LEVel] 2,30
STEP:CURRent[:LEVel] 3,45
STEP:CURRent[:LEVel] 4,60
STEP:CURRent:TIMe 1,1000
; Determine the time interval that the input remains
at each level or point in the STEP before it
advances to the next point. The time is specified in
mill-seconds.
STEP:CURRent:TIMe 2,1500
STEP:CURRent:TIMe 3,2000
STEP:CURRent:TIMe 4,2500
STEP:COUNt 10
; Determine the number of times the STEP is
repeated before it completes. Entering 0 or INFinity
makes the STEP repeat indefinitely. At *RST, the
count is set to 1.
TRIGger:SOURce BUS
INPut ON
STEP:CURRent:STATe AUTO
; the STEP to advance only one point after each
trigger. Triggers that arrive during a dwell delay are
ignored
*TRG
(other statements)
STEP:CURRent:STATe OFF
; Stop the STEP sequencing.
STAT
E
ONC
E
MODe:CURRent
STEP:CURRent[:LEVel] 1,15
; Program the STEP of input values for each
function. The STEP commands take a
comma-separated STEP of arguments.
STEP:CURRent[:LEVel] 2,30
STEP:CURRent[:LEVel] 3,45
STEP:CURRent[:LEVel] 4,60
STEP:CURRent:TIMe 1,1000
; Determine the time interval that the input remains
at each level or point in the STEP before it
advances to the next point. The time is specified in
mill-seconds.
STEP:CURRent:TIMe 2,1500
STEP:CURRent:TIMe 3,2000
STEP:CURRent:TIMe 4,2500
STEP:COUNt 10
; Determine the number of times the STEP is
repeated before it completes. Entering 0 or INFinity
makes the STEP repeat indefinitely. At *RST, the
count is set to 1.
TRIGger:SOURce BUS