Calibrate:point, Calibrate:save, Calibrate:state – AMETEK PLA-PLW Programming Manual User Manual
Page 17

15
10 = range 2 of immediate resistance calibration
11 = range 3 of immediate resistance calibration
12 = range 0 of immediate power calibration
13 = range 1 of immediate power calibration
14 = range 2 of immediate power calibration
15 = range 3 of immediate power calibration
16 = range 0 of transient voltage calibration
17 = range 1 of transient voltage calibration
18 = range 2 of transient voltage calibration
19 = range 3 of transient voltage calibration
20 = range 0 of transient current calibration
21 = range 1 of transient current calibration
22 = range 2 of transient current calibration
23 = range 3 of transient current calibration
24 = range 0 of transient resistance calibration
25 = range 1 of transient resistance calibration
26 = range 2 of transient resistance calibration
27 = range 3 of transient resistance calibration
28 = range 0 of transient power calibration
29 = range 1 of transient power calibration
30 = range 2 of transient power calibration
31 = range 3 of transient power calibration
32 = range 0 of readback voltage calibration
33 = range 1 of readback voltage calibration
34 = range 2 of readback voltage calibration
35 = range 3 of readback voltage calibration
36 = range 0 of readback current calibration
37 = range 1 of readback current calibration
38 = range 2 of readback current calibration
39 = range 3 of readback current calibration
Query Syntax
CALibrate:MODe?
Returned Parameters
CALibrate:POINt
This command can only be used in calibration mode. It is used to set the two calibration points of the
analog current monitor signal.
Command Syntax
CALibrate:POINt
Parameters
st
calibration points
1=2
nd
calibration points
CALibrate:SAVe
This command can only be used in calibration mode. It saves any new calibration constants (after a
current or voltage calibration procedure has been completed) in nonvolatile memory.
Command Syntax
CALibrate:SAVe
Parameters
None
CALibrate:STATe
This command enables and disables calibration mode. The calibration mode must be enabled before
the load will accept any other calibration commands. The query statement returns only the state, not
the password. Whenever the calibration state is changed from enabled to disabled, any new