beautypg.com

Source:]resistance:slew:positive, Source:]resistance:tlevel, rtr, Source:]resistance:[transient:]twidth – AMETEK PLA-PLW Programming Manual User Manual

Page 40

background image

38

Command Syntax

RESistance:SLEW:NEGative

Parameters

slew time value, refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

RESistance:SLEW:NEG?

RESistance:SLEW:NEG? MIN

RES:SLEW:NEG? MAX

Returned Parameters

[SOURce:]RESistance:SLEW:POSitive

This command sets the slew time of the resistance for positive going transitions. MINimum sets the
slew to the fastest possible time. MAXimum sets the slew to the slowest time.

Command Syntax

RESistance:SLEW
RESistance:SLEW:POSitive

Parameters

Refer to model specifications | MIN | MAX

Unit

ms (milli-second)

Query Syntax

RESistance:SLEW?

RESistance:SLEW? MIN

RESistance:SLEW? MAX

RESistance:SLEW:POSitive?

RESistance:SLEW:POSitive? MIN

RESistance:SLEW:POSitive? MAX

Returned Parameters

[SOURce:]RESistance:TLEVel, RTR

Channel-Specific Resistance Command/Query. This command specifies the value of the
programmed resistance level for the TRANsient input when the electronic load is in the CR Mode.
When the Transient Subsystem is on, the electronic load input resistance will switch (under control of
the Transient Subsystem) between the main level and TLEVel at a rate determined by the present
value of SLEW.

In order for the input resistance level to switch, TLEVel must be set to a value greater than the main
level. If TLEVel is set to a value below the main level, no error is generated but switching will not
occur until the main level is subsequently below the value of TLEVel.

Command Syntax

RESistance:TLEVel
RTR

Parameters

resistance transient level setting, refer to individual model specification |
MIN | MAX

Query Syntax

RESistance:TLEVel?

RESistance:TLEVel? MIN

RESistance:TLEV? MAX

RTR?

RTR?

MIN

RTR?

MAX

Returned Parameters

RES:TLEV? returns the transient resistance level.

If the electronic load is not in CR Mode, the level will still be set, even if it
is less than the presently programmed input level.

[SOURce:]RESistance:[TRANsient:]TWIDth

This command sets the pulse width of the transients when the generator is in PULSe mode.

Note: This command will also change the RES:FREQ value.

Command Syntax

RESistance:TRANsient:TWIDth

RESistance:TWIDth