Channel-Specific Current Command/Query. This command specifies the value of the
programmed current level for the TRANsient input when the electronic load is in the CC Mode. When
the Transient Subsystem is on, the electronic load input current will switch (under control of the
Transient Subsystem) between the main level and TLEVel at a rate determined by the present value
of SLEW.

In order for the input current level to switch, TLEVel must be set to a value greater than the main level.
If TLEVel is set to a value below the main level, no error is generated but switching will not occur until
the main level is subsequently below the value of TLEVel.

Command Syntax

CURRent:TLEVel

ITR

Parameters

: Current transient level value. Refer to model specifications for
range | MIN | MAX

Query Syntax

CURRent:TLEVel?
CURRent:TLEVel? MIN

CURRent:TLEVel? MAX

ITR

ITR?

MIN

ITR?

MAX

Returned Parameters

returns the transient current level.

If the electronic load is not in CC Mode, the level will still be set, even if it
is less than the presently programmed input level.

[SOURce:]CURRent:[TRANsient:]TWIDth

This command sets the pulse width of the transients when the generator is in PULSe mode.

Note: This command will also change the CURR:FREQ value.

Command Syntax

CURRent:[TRANsient:]TWIDth

Parameters

Refer to model’s specification | MAX | MIN

Unit

ms

Query Syntax

CURRent:TRANsient:TWIDth?
CURRent:TWIDth?

Returned Parameters

, pulse width time in ms.