Channel-Specific Current Command/Query. This command specifies the value of the
programmed current level for the TRANsient input when the electronic load is in the CC Mode. When
the Transient Subsystem is on, the electronic load input current will switch (under control of the
Transient Subsystem) between the main level and TLEVel at a rate determined by the present value
of SLEW.
In order for the input current level to switch, TLEVel must be set to a value greater than the main level.
If TLEVel is set to a value below the main level, no error is generated but switching will not occur until
the main level is subsequently below the value of TLEVel.
Command Syntax
CURRent:TLEVel
ITR
Parameters
range | MIN | MAX
Query Syntax
CURRent:TLEVel?
CURRent:TLEVel? MIN
CURRent:TLEVel? MAX
ITR
ITR?
MIN
ITR?
MAX
Returned Parameters
If the electronic load is not in CC Mode, the level will still be set, even if it
is less than the presently programmed input level.
[SOURce:]CURRent:[TRANsient:]TWIDth
This command sets the pulse width of the transients when the generator is in PULSe mode.
Note: This command will also change the CURR:FREQ value.
Command Syntax
CURRent:[TRANsient:]TWIDth
Parameters
Refer to model’s specification | MAX | MIN
Unit
ms
Query Syntax
CURRent:TRANsient:TWIDth?
CURRent:TWIDth?