Rigger, Ommands – AMETEK PLA-PLW Programming Manual User Manual
Page 55

51
T
RIGGER
C
OMMANDS
Trigger commands controls the triggering of the electronic load. See also
CURRent:TRIGgered, POWer:TRIGgered, RESistance:TRIGgered, and
VOLTage:TRIGgered in the Input Commands section.
ABORt
This command applies only to trigger functions. It cancels all pending [:LEVel]:TRIG operations (such
as CURR:TRIG) in all operating modes and on all channels. As a result, subsequent triggers have no
effect on the input level. This command resets the WTG bit of the Operation Condition register and
has the same effect on status as the receipt of a trigger. ABORt has no affect on the Transient
Subsystem.
Command Syntax
ABORt
Parameters
None
TRIGger[:IMMediate]
When the trigger system has been initiated, this command generates a trigger signal regardless of the
selected trigger source.
Command Syntax
TRIGger
TRIGger:IMMediate
Parameters
None
TRIGger:DELay
This command sets the time delay between the detection of a trigger signal and the start of any
corresponding trigger action. After the time delay has elapsed, the trigger is implemented.
Command Syntax
TRIG:DEL
Parameters
0 - 65535 | MIN | MAX
Unit
ms
*RST Value
0
Query Syntax
TRIG:DEL?
Returned Parameters
TRIGger:SOURce
This command selects the trigger source.
EXTernal
Selects the electronic load’s trigger input as the trigger source. This
trigger is processed as soon as it is received.
BUS Accepts a GPIB
This selection guarantees that all previous commands are complete
before the trigger occurs.
ETHERNET Selects the Ethernet’s trigger input as the trigger source. This trigger is
processed as soon as it is received.
HOLD Only
the
TRIG:IMM
command will generate a trigger in HOLD mode. All
other trigger commands are ignored.
Command Syntax
TRIGger:SOURce
Parameters
EXTernal | BUS | ETHernet | HOLD