AMETEK PLA-PLW Programming Manual User Manual
Page 62

58
[SOURce:]VOLTage[:TRANsient]:TWIDth
This command sets the pulse width of the transients when the generator is in PULSe mode.
Command Syntax
VOLTage:TRANsient:TWIDth
VOLTage:TWIDth
Parameters
pulse width value, refer to model’s specification | MAX | MIN
Unit
ms
Query Syntax
VOLTage:TRANsient:TWIDth?
VOLTage:TWIDth?
Returned Parameters
Under Voltage Lockout Protection
The programmable under voltage lockout feature allows the user to set the trip point between zero and
full-scale voltage. The input will remain off until the input voltage is greater than the pre-programmed
value.
Continuous Mode
When the V-On is set to continuous mode, the UVL message will appear and the input current will be
forced off each time the voltage drops below the set threshold voltage.
Input-On Mode
Under “input on” mode, the UVL will detect under voltage only once, and will reset each time the
input is turned off. When the voltage is below the UVL threshold, the input current will be forced off
and an “UVL” message will display.
By default, the V-On threshold is set to 1 V and the V-On feature can also prevent turn-on current
spike from occurring. As a precaution, it is recommended to leave the V-On threshold at 1 V or
higher.
˖̂̀̀˴́˷ˍʳ
ʳ
Use " VOLT:VON
Use "VOLT:VON:STAT 0 | 1" to set VON mode. 0 - only work when input ON, 1 - continue