Source:]current:slew:negative, Source:]current:slew[:positive, Source:]current:tlevel, itr – AMETEK PLA-PLW Programming Manual User Manual
Page 22

20
slew time value may be entered, the electronic load selects a slew time that is closest to the
programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time.
Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.
Command Syntax
CURRent:SLEW:BOTH
Parameters
Slew time. Refer to model specifications | MIN | MAX
Unit
ms
Query Syntax
None
[SOURce:]CURRent:SLEW:NEGative
This command sets the slew time of the current for negative going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.
Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.
Command Syntax
CURRent:SLEW:NEGative
Parameters
Slew time. Refer to model specifications | MIN | MAX
Unit
ms
Query Syntax
CURRent:SLEW:NEGative?
CURRent:SLEW:NEGative? MIN
CURRent:SLEW:NEGative? MAX
Returned Parameters
[SOURce:]CURRent:SLEW[:POSitive]
This command sets the slew time of the current for positive going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.
Note: Slew time greater than the maximum value is set to MAXimum. Slew time less than the
minimum value are set to MINimum.
Command Syntax
CURRent:SLEW
CURRent:SLEW:POSitive
Parameters
Unit
ms
Query Syntax
CURRent:SLEW?
CURRent:SLEW? MIN
CURRent:SLEW? MAX
CURRent:SLEW:POSitive?
CURRent:SLEW:POSitive? MIN
CURRent:SLEW:POSitive? MAX
Returned Parameters
[SOURce:]CURRent:TLEVel, ITR