AMETEK PLA-PLW Programming Manual User Manual
Page 64

60
I
NPUT
P
OWER
All models have a programmable power function. The command to program the current
is:
POWer
where
Maximum Power
The maximum input power that can be programmed can be queried with:
POWer? MAX
Overpower Protection
The electronic load can also be programmed to turn off its input if the power protection
level is reached. This protection feature is implemented the following command:
POWer:PROTection
POWer:PROTection:STATe ON | OFF
Underpower Protection
The electronic load can also be programmed to turn off its input if its input current is
lower than the under-power protection level. This protection feature is implemented the
following command:
POWer:PROTection:UNDer
POWer:PROTection:UNDer:STATe ON | OFF
I
NPUT
R
ESISTANCE
The input resistance is controlled with the RESistance command. For example, to set
the input resistance to 25 ohms, use:
RESistance 25
Maximum Resistance
The maximum input resistance that can be programmed can be queried with:
RESistance? MAX
I
NPUT
V
OLTAGE
The input voltage is controlled with the VOLTage command. For example, to set the
input voltage to 25 volts, use:
VOLTage 25
Maximum Voltage
The maximum input voltage that can be programmed can be queried with: