Returned Parameters

: “CURR?” and “CURR:TRIG?” return the presently programmed
current levels. After a trigger signal or “ABORt”, “CURR:TRIG?” returns
the same value as “CURR?”.

“CURR?

MAX”, “CURR? MIN”, “CURR:TRIG? MAX” and

“CURR:TRIG?” MIN return the maximum and minimum programmable
LEVel values.

[SOURce:]CURRent:[TRANsient:]DUTY

This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.

Command Syntax

CURRent:[TRANient:]DUTY

Parameters

: Duty cycle value. 1 – 100.

Unit

Percentage (%)

Query Syntax

CURRent:DUTY?

CURRent:DUTY?

MIN

CURRent:DUTY?

MAX

CURRent:TRANsient:DUTY?

CURRent:TRANsient:DUTY? MIN

CURRent:TRANsient:DUTY? MAX

Returned Parameters

, duty cycle value in percentage

[SOURce:]CURRent:[TRANsient:]FREQuency

This command sets the frequency of the transients when the generator is in CONTinuous mode.

Command Syntax

CURRent:[TRANsient:]FREQuency

Parameters

: Frequency value, refer to model specifications for range.

Unit

Hertz

Query Syntax

CURRent:FREQuency?

CURRent:FREQuency?

MIN

CURRent:FREQuency?

MAX

CURRent:TRANsient:FREQuency?

CURRent:TRANsient:FREQuency? MIN

CURRent:TRANsient:FREQuency? MAX

Returned Parameters

[SOURce:]CURRent:PROTection[:LEVel]
[SOURce:]CURRent:PROTection:DELay
[SOURce:]CURRent:PROTection:OVER[:LEVel]
[SOURce:]CURRent:PROTection:OVER:DELay
[SOURce:]CURRent:PROTection:STATe
[SOURce:]CURRent:PROTection:UNDer[:LEVel]
[SOURce:]CURRent:PROTection:UNDer:DELay
[SOURce:]CURRent:PROTection:UNDer:STATe

Channel-Specific Current Limiting Command/Query. This command sets the over-current
protection limit or the under-current protection limit to the input current that the electronic load will sink.
When the input current reaches the over-current protection limit or falls below the under-current
protection limit for the specified delay period, the input of the electronic load is shut off and draws no
current.

The INPut:PROTection:CLEar command (or front panel key) re-enables the input current. The trigger
activated current functions (CURR[:LEV]:TRIG and CURR:TLEV) automatically keep track of