Source:]current:[transient:]duty, Source:]current:[transient:]frequency – AMETEK PLA-PLW Programming Manual User Manual
Page 20

18
Returned Parameters
current levels. After a trigger signal or “ABORt”, “CURR:TRIG?” returns
the same value as “CURR?”.
“CURR?
MAX”, “CURR? MIN”, “CURR:TRIG? MAX” and
“CURR:TRIG?” MIN return the maximum and minimum programmable
LEVel values.
[SOURce:]CURRent:[TRANsient:]DUTY
This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.
Command Syntax
CURRent:[TRANient:]DUTY
Parameters
Unit
Percentage (%)
Query Syntax
CURRent:DUTY?
CURRent:DUTY?
MIN
CURRent:DUTY?
MAX
CURRent:TRANsient:DUTY?
CURRent:TRANsient:DUTY? MIN
CURRent:TRANsient:DUTY? MAX
Returned Parameters
[SOURce:]CURRent:[TRANsient:]FREQuency
This command sets the frequency of the transients when the generator is in CONTinuous mode.
Command Syntax
CURRent:[TRANsient:]FREQuency
Parameters
Unit
Hertz
Query Syntax
CURRent:FREQuency?
CURRent:FREQuency?
MIN
CURRent:FREQuency?
MAX
CURRent:TRANsient:FREQuency?
CURRent:TRANsient:FREQuency? MIN
CURRent:TRANsient:FREQuency? MAX
Returned Parameters
[SOURce:]CURRent:PROTection[:LEVel]
[SOURce:]CURRent:PROTection:DELay
[SOURce:]CURRent:PROTection:OVER[:LEVel]
[SOURce:]CURRent:PROTection:OVER:DELay
[SOURce:]CURRent:PROTection:STATe
[SOURce:]CURRent:PROTection:UNDer[:LEVel]
[SOURce:]CURRent:PROTection:UNDer:DELay
[SOURce:]CURRent:PROTection:UNDer:STATe
Channel-Specific Current Limiting Command/Query. This command sets the over-current
protection limit or the under-current protection limit to the input current that the electronic load will sink.
When the input current reaches the over-current protection limit or falls below the under-current
protection limit for the specified delay period, the input of the electronic load is shut off and draws no
current.
The INPut:PROTection:CLEar command (or front panel key) re-enables the input current. The trigger
activated current functions (CURR[:LEV]:TRIG and CURR:TLEV) automatically keep track of