System:cross:mode:state, System:cross:range:state, System:eos – AMETEK PLA-PLW Programming Manual User Manual
Page 49: System:error

45
Returned Parameters
SYSTem:CROSs:MODe:STATe
This command sets or queries the cross mode state.
Command Syntax
SYSTem:CROSs:MODe:STATe
Parameters
0 = Input Off
1 = Input Remain
Query Syntax
SYSTem:CROSs:MODe:STATe?
Parameters
None
Returned Parameters
SYSTem:CROSs:RANGe:STATe
This command sets or queries the cross range state.
Command Syntax
SYSTem:CROSs:RANGe:STATe
Parameters
0 = Input Off
1 = Input Remain
Query Syntax
SYSTem:CROSs:RANGe:STATe?
Parameters
None
Returned Parameters
SYSTem:EOS
This command sets or queries the value of the end-of-string(EOS) code for all interface (RS-232
interface, GPIB interface and Ethernet interface).
Command Syntax
SYSTem:EOS
Parameters
0 = NULL(no EOS)
1 = CR
2 = LF
3 = CR/LF
Query Syntax
SYSTem:EOS?
Parameters
None
Returned Parameters
SYSTem:ERRor?
This query returns the next error message response string from the remote programming error queue.
The queue is a FIFO (first-in, first-out) buffer that stores maximum 9 errors as they occur. As it is read,
each error is removed from the queue. When all errors have been read, the query returns “0,No
error”. Please refer to the following table for detailed error code, and error message.
Query Syntax
SYSTem:ERRor?
Parameters
None
Returned Parameters
“SYSTem:ERRor?” returned code and error message table: