Root-level commands are those that are specific to the family of electronic loads. The
commands are grouped as either channel-specific or channel-independent commands

C

ALIBRATION

C

OMMANDS

Calibration commands let you:

Ϯ Enable and disable the calibration mode
Ϯ Change the calibration password
Ϯ Calibrate the input functions, current monitor offset, and store new calibration

constants in nonvolatile memory.

CALibrate:MEASure:HIGH

This command can only be used in calibration mode. It is used to set the count of 2nd
calibration point.

Command Syntax

CALibrate:MEASure:HIGH

Parameters

count of calibration point

Query Syntax

CALibrate:MEASure:HIGH?

Returned Parameters

CALibrate:MEASure:LOW

This command can only be used in calibration mode. It is used to set the count of 1st
calibration point.

Command Syntax

CALibrate:MEASure:LOW

Parameters

count of calibration point

Query Syntax

CALibrate:MEASure:LOW?

Returned Parameters

CALibrate:MODe

This command can only be used in calibration mode. It is used to set the calibration mode.

Command Syntax

CALibrate:MODe

Parameters

calibration mode

0 = range 0 of immediate voltage calibration

1 = range 1 of immediate voltage calibration

2 = range 2 of immediate voltage calibration

3 = range 3 of immediate voltage calibration

4 = range 0 of immediate current calibration

5 = range 1 of immediate current calibration

6 = range 2 of immediate current calibration

7 = range 3 of immediate current calibration

8 = range 0 of immediate resistance calibration

9 = range 1 of immediate resistance calibration