Root-level commands are those that are specific to the family of electronic loads. The
commands are grouped as either channel-specific or channel-independent commands
C
ALIBRATION
C
OMMANDS
Calibration commands let you:
Ϯ Enable and disable the calibration mode
Ϯ Change the calibration password
Ϯ Calibrate the input functions, current monitor offset, and store new calibration
constants in nonvolatile memory.
CALibrate:MEASure:HIGH
This command can only be used in calibration mode. It is used to set the count of 2nd
calibration point.
Command Syntax
CALibrate:MEASure:HIGH
Parameters
Query Syntax
CALibrate:MEASure:HIGH?
Returned Parameters
CALibrate:MEASure:LOW
This command can only be used in calibration mode. It is used to set the count of 1st
calibration point.
Command Syntax
CALibrate:MEASure:LOW
Parameters
Query Syntax
CALibrate:MEASure:LOW?
Returned Parameters
CALibrate:MODe
This command can only be used in calibration mode. It is used to set the calibration mode.
Command Syntax
CALibrate:MODe
Parameters
0 = range 0 of immediate voltage calibration
1 = range 1 of immediate voltage calibration
2 = range 2 of immediate voltage calibration
3 = range 3 of immediate voltage calibration
4 = range 0 of immediate current calibration
5 = range 1 of immediate current calibration
6 = range 2 of immediate current calibration
7 = range 3 of immediate current calibration
8 = range 0 of immediate resistance calibration
9 = range 1 of immediate resistance calibration