Source:]voltage:slew:both – AMETEK PLA-PLW Programming Manual User Manual
Page 60
56
incoming triggers while the input is shut down and will respond to the trigger as soon as the protection
fault is cleared.
The :PROTection:DELay command specifies the time that the input voltage may equal or exceed
VOLTage:PROTection[:LEVel] or equal or fall below the VOLTage:PROTection:UNDer[:LEVel] before
the soft circuit breaker is actuated. The PROTection:STATe command enables or disables the soft
circuit breaker function.
Note : If the soft circuit breaker function causes the input to shut down, it will not affect INP[STATe]. If
INP:STAT is programmed ON, it will remain so even after the VOLTage:PROT has turned the
electronic load off.
Command Syntax
VOLTage:PROTection
protection
limit.
VOLTage:PROTection:DELay
Set time that voltage may be at or
above :LEVel before input is turned off.
VOLTage:PROTection:STATe OFF|0
Disable over protection function.
VOLTage:PROTection:STATe ON|1
Enable over protection function.
VOLTage:PROTection:UNDer
Set immediate under-voltage protection
limit.
VOLTage:PROTection:UNDer:STATe OFF|0
Disable under protection function.
VOLTage:PROTection:UNDer:STATe ON|1
Enable under protection function.
Parameters
Refer individual model specification | MIN | MAX
Query Syntax
VOLTage:PROTection?
VOLTage:PROTection? MIN
VOLTage:PROTection? MAX
VOLTage:PROTection:OVER?
VOLTage:PROTection:OVER? MIN
VOLT:PROTection:OVER? MAX
VOLTage:PROTection:DELay?
VOLTage:PROTection:DELay? MIN
VOLTage:PROTection:DELay? MAX
VOLTage:PROTection:OVER:DELay?
VOLTage:PROTection:OVER:DELay? MIN
VOLTage:PROTection:OVER:DELay? MAX
VOLTage:PROTection:UNDer:DELay?
VOLTage:PROTection:UNDer:DELay? MIN
VOLTage:PROTection:UNDer:DELay? MAX
VOLTage:PROTection:STATe?
VOLTage:PROTection:UNDer?
VOLTage:PROTection:UNDer? MIN
VOLTage:PROTection:UNDer? MAX
VOLTage:PROTection:UNDer:STATe?
[SOURce:]VOLTage:SLEW:BOTH
This command sets the slew time for all programmed changes in the input voltage level of the
electronic load. This command programs both positive and negative going slew time. Although any
slew time value may be entered, the electronic load selects a slew time that is closest to the
programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time. Slew times less than the minimum value are set to MINimum. Slew times greater
than the maximum value are set to MAXimum.
Command Syntax
VOLTage:SLEW:BOTH