These commands program the transient generator of the electronic load. The transient
generator programs a second (transient) level at which the electronic load can operate
without changing the original programmed settings.
See also [SOURce:]CURRent:TLEVel, [SOURce:]RESistance:TLEVel, and
[SOURce:]VOLTage:TLEVel in the Input Commands section.
[SOURce:]TRANsient
This command turns the transient generator on or off.
Command Syntax
TRANsient
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Query Syntax
TRANsient?
Returned Parameters
[SOURce:]TR:MODe
[SOURce:]TRANsient:MODe
This command selects the operating mode of the transient generator as follows.
TRIGger
The transient switch to trigger levels upon receipt of a trigger.
CONTinuous The
transient generator puts out a continuous pulse stream.
TOGGle
The
transient generator toggles between two levels upon receipt of a
trigger.
PULSe
The
transient generator puts out a single pulse upon receipt of a trigger.
Command Syntax
TR:MODe
TRANsient:MODe
Parameters
0:TRIGger
1:CONTinuous
2:TOGGle
3:PULSe
4:STEP
5:AUTO
6:ONCE
Query Syntax
TR:MODe?
TRANsient:MODe?
Returned Parameters