Source:]current:slew:both – AMETEK PLA-PLW Programming Manual User Manual
Page 21

19
incoming triggers while the input is shut down and will respond to the trigger as soon as the protection
fault is cleared.
The :PROTection:DELay command specifies the time that the input current may equal or exceed
CURRent:PROTection[:LEVel] or equal or fall below the CURRent:PROTection:UNDer[:LEVel] before
the soft circuit breaker is actuated. The PROTection:STATe command enables or disables the soft
circuit breaker function.
Note : If the soft circuit breaker function causes the input to shut down, it will not affect INP[STATe]. If
INP:STAT is programmed ON, it will remain so even after the CURR:PROT has turned the electronic
load off.
Command Syntax
CURRent:PROTection
protection
limit.
CURRent:PROTection:OVER
Set immediate over-current
protection
limit.
CURRent:PROTection:DELay
Set time that current may be at or above
limit before input is turned off.
CURRent:PROTection:OVER:DELay
Set time that current may be at or above
limit before input is turned off.
CURRent:PROTection:UNDer:DELay
Set time that current may be at or below
limit before input is turned off.
CURRent:PROTection:STATe OFF|0
Disable over protection function.
CURRent:PROTection:STATe ON|1
Enable over protection function.
CURRent:PROTection:UNDer
Set immediate under-current protection
limit.
CURRent:PROTection:UNDer:STATe OFF|0
Disable under current protection function.
CURRent:PROTection:UNDer:STATe ON|1
Enable under current protection function.
Parameters
Refer individual model specification for MIN | MAX value.
Query Syntax
CURRent:PROTection?
CURRent:PROTection? MIN CURRent:PROTection? MAX
CURRent:PROTection:OVER?
CURRent:PROTection:OVER? MIN
CURRent:PROTection:OVER? MAX
CURRent:PROTection:DELay?
CURRent:PROTection:DELay? MIN
CURRent:PROTection:DELay? MAX
CURRent:PROTection:OVER:DELay?
CURRent:PROTection:OVER:DELay? MIN
CURRent:PROTection:OVER:DELay?MAX
CURRent:PROTection:UNDer:DELay?
CURRent:PROTection:UNDer:DELay? MIN
CURRent:PROTection:UNDer:DELay?MAX
CURRent:PROTection:STATe?
CURRent:PROTection:UNDer?
CURRent:PROTection:UNDer?
MIN
CURRent:PROTection:UNDer?
MAX
CURRent:PROTection:UNDer:STATe?
Returned Parameters
Depending on the actual query.
[SOURce:]CURRent:SLEW:BOTH
This command sets the slew time for all programmed changes in the input current level of the
electronic load. This command programs both positive and negative going slew time. Although any