Easurement, Ommands – AMETEK PLA-PLW Programming Manual User Manual
Page 30

28
M
EASUREMENT
C
OMMANDS
MEASure:CURRent[:DC]?
MEASure:POWer[:DC]?
MEASure:RESistance[:DC]?
MEASure:VOLTage[:DC]?
MEASure:VOLTage:INPut?
This function consists of queries that return the current, power, resistance ,and voltage at the
input of the electronic load.
Query Syntax
MEASure:CURRent?
electronic
load input current
MEASure:POWer? Computed electronic load input power
MEASure:RESistance? Computed electronic load input resiatance
MEASure:VOLTage?
electronic
load input voltage
MEASure:VOLT:INPut? electronic
load input voltage at input end
Returned parameters
MEASure:DELay
This command is to sets or queries the delay time to obtain read back value for “MEAS:CURR?”,
“MEAS:POW?”, “MEAS:RES?” and “MEAS:VOLT?” queries.
Command Syntax
MEASure:DELay
Parameters
0 t0 2000 | MIN | MAX
Unit
milli-second
Query Syntax
MEASure:DELay?
MEASure:DELay? MIN
MEASure:DELay? MAX
Returned Parameters
MEASure:SENSe
This command sets or queries the measurement sense (Local / Remote ).
Note: When input voltage is greater than 30 V, this command will cause an execution error.
Command Syntax
MEASure:SENSe
Parameters
0 | 1 | LOCal(0) | REMote(1)
Query Syntax
MEASure:SENSe?
Returned Parameters
1=REMote