programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time. Slew times less than the minimum value are set to MINimum. Slew time greater than
the maximum value are set to MAXimum.

Command Syntax

POW:SLEW:BOTH

Parameters

Slew time, refer to model specifications | MIN | MAX

Unit

ms

[SOURce:]POWer:SLEW:NEGative

This command sets the slew time of the power for negative going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.

Command Syntax

POWer:SLEW:NEGative

Parameters

Slew time, refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

POWer:SLEW:NEGative?

POWer:SLEW:NEGative? MIN

POWer:SLEW:NEGative? MAX

Returned Parameters

slew time in ms

[SOURce:]POWer:SLEW[:POSitive]

This command sets the slew time of the power for positive going transitions. MINimum sets the slew
to the fastest possible time. MAXimum sets the slew to the slowest time.

Command Syntax

POWer:SLEW
POWer:SLEW:POSitive

Parameters

slew time, refer to model specifications | MIN | MAX

Unit

ms

Query Syntax

POWer:SLEW?

POWer:SLEW?

MIN

POWer:SLEW? MAX

POWer:SLEW:POSitive?

POWer:SLEW:POSitive? MIN

POWer:SLEW:POSitive? MAX

Returned Parameters

slew time in ms

[SOURce:]POWer:TLEVel, PTR

Channel-Specific Power Command/Query. This command specifies the value of the programmed
power level for the TRANsient input when the electronic load is in the CP Mode. When the Transient
Subsystem is on, the electronic load input power will switch (under control of the Transient Subsystem)
between the main level and TLEVel at a rate determined by the present value of SLEW.

In order for the input power level to switch, TLEVel must be set to a value greater than the main level.
If TLEVel is set to a value below the main level, no error is generated but switching will not occur until
the main level is subsequently below the value of TLEVel.

Command Syntax

POWer:TLEVel

PTR

Parameters

Power transient level value, refer to model specifications for range.

Query Syntax

POWer:TLEVel?

POWer:TLEVel? MIN

POWer:TLEVel? MAX

PTR?

PTR?

MIN

PTR?

MAX

Returned Parameters

“POW:TLEV?” returns the transient power level.