Calibrate:transient:offset, Calibrate:value:current, Calibrate:value:voltage – AMETEK PLA-PLW Programming Manual User Manual
Page 18

16
calibration constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax
CALibrate:STATe
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Query Syntax
CALibrate:STATe?
Returned Parameters
CALibrate:TRANsient:OFFSet
This command can only be used in calibration mode. It is used to set the DAC count of
immediate level when enter transient calibration mode.
Command Syntax
CALibrate:TRANsient:OFFSet
Parameters
Query Syntax
CALibrate:TRANsient:OFFSet?
Returned Parameters
CALibrate:VALue:CURRent
This command is only used in calibration mode. It enters a calibration current that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.
Command Syntax
CALibrate:VALue:CURRent
Parameters
CALibrate:VALue:VOLTage
This command is only used in calibration mode. It enters a calibration voltage that you obtain by
reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value
being entered. These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored.
Command Syntax
CALibrate:VALue:VOLTage
Parameters