Source:]resistance:[transient:]duty, Source:]resistance:[transient:]frequency, Source:]resistance:slew:both – AMETEK PLA-PLW Programming Manual User Manual

Page 39: Source:]resistance:slew:negative

background image

37

Returned Parameters

RES? and RES:TRIG?
return the presently programmed resistance levels. After a trigger or
ABORt, RESistance:TRIG? returns the same value as RESistance? .

RESistance? MAX, RESistance? MIN, RESistance:TRIG? MAX and
RESistance:TRIG? MIN return the maximum and minimum
programmable LEVel and TLEVel values.

[SOURce:]RESistance:[TRANsient:]DUTY

This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.

Command Syntax

RESistance:DUTY

RES:TRAN:DUTY

Parameters

Duty cycle value, 1 - 100

Unit

Percentage

Query Syntax

RESistance:DUTY?

RESistance:DUTY? MIN

RESistance:DUTY? MAX

RESistance:TRANsient:DUTY?

RESistance:TRANsient:DUTY? MIN

RESistance:TRANsient:DUTY? MAX

Returned Parameters

duty cycle in percentage

[SOURce:]RESistance:[TRANsient:]FREQuency

This command sets the frequency of the transients when the generator is in CONTinuous mode.

Command Syntax

RESistance: FREQuency
RESistance:TRANsient:FREQuency

Parameters

Frequency value, refer to model specifications | MAX | MIN

Unit

Hertz

Query Syntax

RESistance:FREQuency?

RESistance:FREQuency? MIN

RESistance:FREQuency? MAX

RESistance:TRANsient:FREQuency?

RESistance:TRANsient:FREQuency? MIN

RESistance:TRANsient:FREQuency? MAX

Returned Parameters

frequency value.

[SOURce:]RESistance:SLEW:BOTH

This command sets the slew time for all programmed changes in the input resistance level of the
electronic load. This command programs both positive and negative going slew time. Although any
slew time value may be entered, the electronic load selects a slew time that is closest to the
programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time. Slew times less than the minimum value are set to MINimum. Slew times greater
than the maximum value are set to MAXimum.

Command Syntax

RESistance:SLEW:BOTH

Parameters

Slew time value, refer to model specifications | MIN | MAX

Unit

ms

[SOURce:]RESistance:SLEW:NEGative

This command sets the slew time of the resistance for negative going transitions. MINimum sets the
slew to the fastest possible time. MAXimum sets the slew to the slowest time.