37
Returned Parameters
return the presently programmed resistance levels. After a trigger or
ABORt, RESistance:TRIG? returns the same value as RESistance? .
RESistance? MAX, RESistance? MIN, RESistance:TRIG? MAX and
RESistance:TRIG? MIN return the maximum and minimum
programmable LEVel and TLEVel values.
[SOURce:]RESistance:[TRANsient:]DUTY
This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.
Command Syntax
RESistance:DUTY
RES:TRAN:DUTY
Parameters
Duty cycle value, 1 - 100
Unit
Percentage
Query Syntax
RESistance:DUTY?
RESistance:DUTY? MIN
RESistance:DUTY? MAX
RESistance:TRANsient:DUTY?
RESistance:TRANsient:DUTY? MIN
RESistance:TRANsient:DUTY? MAX
Returned Parameters
[SOURce:]RESistance:[TRANsient:]FREQuency
This command sets the frequency of the transients when the generator is in CONTinuous mode.
Command Syntax
RESistance: FREQuency
RESistance:TRANsient:FREQuency
Parameters
Frequency value, refer to model specifications | MAX | MIN
Unit
Hertz
Query Syntax
RESistance:FREQuency?
RESistance:FREQuency? MIN
RESistance:FREQuency? MAX
RESistance:TRANsient:FREQuency?
RESistance:TRANsient:FREQuency? MIN
RESistance:TRANsient:FREQuency? MAX
Returned Parameters
[SOURce:]RESistance:SLEW:BOTH
This command sets the slew time for all programmed changes in the input resistance level of the
electronic load. This command programs both positive and negative going slew time. Although any
slew time value may be entered, the electronic load selects a slew time that is closest to the
programmed value. MINimum sets the slew to the fastest possible time. MAXimum sets the slew to
the slowest time. Slew times less than the minimum value are set to MINimum. Slew times greater
than the maximum value are set to MAXimum.
Command Syntax
RESistance:SLEW:BOTH
Parameters
Slew time value, refer to model specifications | MIN | MAX
Unit
ms
[SOURce:]RESistance:SLEW:NEGative
This command sets the slew time of the resistance for negative going transitions. MINimum sets the
slew to the fastest possible time. MAXimum sets the slew to the slowest time.