Source:]voltage:[transient:]duty, Source:]voltage:[transient:]frequency – AMETEK PLA-PLW Programming Manual User Manual
Page 59

55
return the presently programmed voltage levels. After a trigger or ABORt,
VOLTage:TRIG? returns the same value as VOLTage? .
VOLTage? MAX, VOLTage? MIN, VOLTage:TRIG? MAX and
VOLTage:TRIG? MIN return the maximum and minimum programmable
LEVel and TLEVel values.
[SOURce:]VOLTage:[TRANsient:]DUTY
This command sets the duty cycle of each of the transients when the generator is in CONTinuous
mode.
Command Syntax
VOLTage:DUTY
VOLTage:TRANsient:DUTY
Parameters
Duty cycle value, 1 - 100
Unit
Percentage
Query Syntax
VOLTage:DUTY?
VOLTage:DUTY? MIN
VOLTage:DUTY? MAX
VOLTage:TRANsient:DUTY?
VOLTage:TRANsient:DUTY? MIN
VOLTage:TRANsient:DUTY? MAX
Returned Parameters
[SOURce:]VOLTage:[TRANsient:]FREQuency
This command sets the frequency of the transients when the generator is in CONTinuous mode.
Command Syntax
VOLTage:FREQuency
VOLTage:TRANsient:FREQuency
Parameters
Frequency value, refer to model specifications | MIN | MAX
Unit
Hertz
Query Syntax
VOLTage:FREQuency?
VOLTage: FREQuency? MIN
VOLTage:FREQuency? MAX
VOLTage:TRANsient:FREQuency?
VOLTage:TRANsient:FREQuency? MIN
VOLTage:TRANsient:FREQuency? MAX
Returned Parameters
[SOURce:]VOLTage:PROTection[:LEVel]
[SOURce:]VOLTage:PROTection:OVER[:LEVel]
[SOURce:]VOLTage:PROTection:DELay
[SOURce:]VOLTage:PROTection:OVER:DELay
[SOURce:]VOLTage:PROTection:UNDer:DELay
[SOURce:]VOLTage:PROTection:STATe
[SOURce:]VOLTage:PROTection:UNDer[:LEVel]
[SOURce:]VOLTage:PROTection:UNDer:STATe
Channel-Specific Voltage Limiting Command/Query. This command sets the over-voltage
protection limit or the under-voltage protect limit to the input voltage that the electronic load will sink.
When the input voltage reaches the protection limit for the specified delay period, the input of the
electronic load is shut off and draws no voltage.
The INPut:PROTection:CLEar command (or front panel key) re-enables the input voltage. The trigger
activated voltage functions (VOLTage[:LEV]:TRIG and VOLTage:TLEV) automatically keep track of