STATus:QUEStionable?

This query returns the value of the Questionable Event register. The Event register is a read-only
register that holds (latches) all events that pass into it. Reading the Questionable Event register clears
it. This command is not channel specific, it applies to the entire mainframe.

Query Syntax

STATus:QUEStionable?

Parameters

None

Returned Parameters

(register value)

STATus:QUEStionable:CONDition?

This query returns the value of the Questionable Condition register. That is a read-only register that
holds the real-time (unlatched) questionable status of the electronic load. This command is not
channel specific, it applies to the entire mainframe.

Query Syntax

STATus:QUEStionable:CONDition?

Parameters

None

Returned Parameters

(register value)

STATus:QUEStionable:ENABle

This command sets or reads the value of the Questionable Enable register. This register is a mask for
enabling specific bits from the Questionable Event register to set the questionable summary (QUES)
bit of the Status Byte register. This bit (bit 3) is the logical OR of all the Questionable Event register
bits that are enabled by the Questionable Status Enable register. This command is not channel
specific, it applies to the entire mainframe.

Command Syntax

STATus:QUEStionable:ENABle

Parameters

0 to 32767 | MAXimum | MINimum

Default Value

0

Query Syntax

STATus:QUEStionable:ENABle?

Returned Parameters

(register value)