Testing Fibers in Advanced Mode

OTDR

79

Setting Pass/Fail Thresholds

The following table provides the default, minimum and maximum
thresholds.

Once the thresholds are set, the application will be able to perform
Pass/Fail tests to determine the status of the various events (pass, warning,
fail).

The Pass/Fail test is performed on two occasions:

³

when analyzing or reanalyzing a trace

³

when you open a trace file

By default, when the thresholds are set, the application displays symbols in
the Result tab to identify the events’ status. Values that are greater than the
predefined fail thresholds are displayed in white on a red background in
the events table. Values that are greater than the predefined warning
thresholds are displayed in black on a yellow background.

You can also set the application to display pass/fail messages when the
Pass/Fail test is performed (see Displaying or Hiding Pass/Fail Messages on
page 114).

Test

Default

Minimum

Maximum

Splice loss (dB)

0.500

0.015

5.000

Connector loss (dB)

1.000

0.015

5.000

Reflectance (dB)

–40.00

–80.00

0.00

Fiber section attenuation (dB/km)

0.40

0.00

5.000

Span loss (dB)

45.000

0.000

45.000

Span length (km)

0.00

0.0000

300.0000

Span ORL (dB)

15.00

15.00

40.000