beautypg.com

EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 292

background image

Troubleshooting

278

OTDR

Error Messages

Communication test

with the module has

failed.

The module is not able to

perform the commands

properly.

Contact EXFO.

Unable to read

current version of the

calibration EEPROM.

The software version installed

on your FTB-500 is too old for

the module currently being

used.

Update ToolBox software

version (refer to the FTB-400
Universal Test System
user

guide).Update the OTDR

software version (refer to the

Update Manager online help).

Module memory

error.

Impossible to access the

memory where data points are

stored.

Contact EXFO.

Unable to adjust the

amplification chain’s

offset.

Impossible to set an internal

component (ADC) to the

appropriate position when the

photodetector is not connected.
The module is probably

defective.

Contact EXFO.

Unable to adjust

offset with APD

connected.

Impossible to set an internal

component (ADC) to the

appropriate position when the

photodetector is connected.
Light is suddenly detected in the

module even though no sign of a

live fiber was detected at the

beginning of the acquisition.

³

Ensure that no live fiber is

connected to the OTDR

port.

³

Stop any acquisition that
could be underway,

disconnect the fiber from

the OTDR port and close

the connector’s cap to

ensure no light will reach

the port. Start a new

acquisition.

If the problem persists, contact

EXFO.

Error Message

Possible Cause

Solution