background image

Customizing the Application

122

OTDR

Defining the Number of Digits Displayed after the Decimal Point

Defining the Number of Digits Displayed after

the Decimal Point

You can set the number of digits that will be displayed after the decimal
point for the following values:

³

Span loss

³

Reflectance

³

Section attenuation

³

Span length

³

Span ORL

This will affect the way values are displayed and, possibly, the status of the
results (pass, warning or fail).

The following table indicates what would happen with a particular fiber
section having an attenuation value of 0.5523.

Note: The displayed values are rounded, not truncated.

Note: This function is available in Advanced mode only.

Value

Number of digits Displayed value

Warning

threshold

Result status

0.5523

3

0.552

0.550

Warning

0.5523

2

0.55

0.55

Pass