17 Troubleshooting

Solving Common Problems

Problem

Cause

Solution

New module is not

working.

The software version

installed on your

FTB-500 is too old for

the module currently

being used.

Update the OTDR software version

using the CD that came with your

new module (refer to the Update

Manager online help).

The application does

not use your custom

thresholds.

The thresholds have

been defined on the

wrong wavelength.

Ensure that the desired wavelength is

selected before saving the new

thresholds or apply the new

thresholds to all wavelengths. For

more information, see Setting
Pass/Fail Thresholds
on page 78.

The application displays

a message indicating

that a “Non-resolved

fiber end” event has

been found.

The fiber under test is

too long.

Ensure that the fiber under test is

shorter than the maximum length the

OTDR can measure.

In multimode fiber

testing, launch level

remains out of the

launch window (light

green rectangle) even

after cleaning and

verifying connection.

Wrong fiber type

selected.

³

If you are testing C fiber, from the
Auto or Advanced main window,

select MM 50 μm.

³

If you are testing D fiber, from the
Auto or Advanced main window,

select MM 62.5 μm.