beautypg.com

Mportant – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 270

background image

Preparing for Automation or Remote Control

256

OTDR

³

Auto Setting Results: Acquisition values suggested by the

application to get the best possible results. If you want to use this
feature, remember to set the OTDR’s acquisition mode to ACQuisition
first.

³

Status: Current state of your OTDR (initialization in progress,

ready, etc.) and error messages.

I

MPORTANT

The OTDR parameters are NOT automatically set to the suggested
values. You must set them yourself using the appropriate SCPI
commands.

6

See :FETCh[1..n]:ASETting:RANGe?

on page 399

See :FETCh[1..n]:ASETting:PULSe?

on page 398

See :FETCh[1..n]:ASETting:DURation?

on page 397

7

See :INITiate[1..n]:STATe? on page 411

:ERRor[1..n]? on page 395