beautypg.com

Viewing and modifying current trace settings – EXFO FTB-7000 OTDR Series for FTB-500 User Manual

Page 167

background image

Analyzing Traces and Events

OTDR

153

Viewing and Modifying Current Trace Settings

Viewing and Modifying Current Trace Settings

You can view the trace parameters and modify them at your convenience.

Note: Parameter modification is only possible in Advanced mode and in Auto

mode (if you selected the Enable Editing for Current Trace Settings in the
Mode tab). For more information on the activation and deactivation of this
feature, see Testing Fibers in Auto Mode
on page 57.

Two groups of parameters can be changed:

³

Fiber settings: index of refraction (IOR) also known as group index,
Rayleigh backscatter (RBS) coefficient, and helix factor.

³

Analysis detection thresholds: for splice loss, reflectance, and
end-of-fiber detection.

Modifications you make are only applied to the current trace (that is, to a
particular wavelength), not to all traces.

These modifications alter the displayed traces. These settings will also be
used when you reanalyze the trace.

The application will only prompt you to reanalyze the trace if you modify
the RBS coefficient (no analysis necessary when you modify the IOR or
helix factor). If you want to modify the parameters that will be used for
future acquisitions, see Setting the IOR, RBS Coefficient, and Helix Factor
on page 69 and
Setting the Analysis Detection Thresholds on page 167.